Notifications
You must be signed in to change notification settings - Fork 2
Astrium Chopper
John Holt edited this page Jan 21, 2019
·
18 revisions
Wiki > The Backend System > Specific Device IOC > Astrium Chopper
Chopper 1 (NCS016)
If the control program on the chopper PC is restarted for any reason, a reset will need to be performed before the chopper can be calibrated (which is also needed after a program restart). This is done with a physical toggle switch on the monitoring system crate in the top of the left-hand cabinet. (Fault LED 8 on the same card will be illuminated showing the error).
This cabinet is normally locked and so an instrument scientist will need to be contacted to gain access.
Chopper 5 (NCS017)
It is believed that the reset is not required on this system.