Something Has Gone Wrong with my Experimental Run
This is from a problem see on IRIS.
- Get the script and configuration.
In this case the important bits are:
def set_sample_temp(temp, high, low, tlccr_offset): g.cset(Sample=temp, highlimit=high, lowlimit=low, runcontrol=True) g.cset(TLCCR=temp+tlccr_offset, runcontrol=False) def next_sample(temp, uamps): g.change(title="Fast Frozen Liver Cells transplant + Cryoprotectant + algae "+str(temp)+"K PG002") g.begin() g.waitfor(uamps=uamps) g.end() for i in range(47): temp += 5 set_sample_temp(temp, temp+2, temp-2,-5) next_sample(temp, 20)
- Find the PVs and plot them in log plotter
- Open IBEX GUI. Switch to the instrument in question
- If the config is loaded right click on PV and display block history (otherwise do this and edit)
- Look for the time in question (zoom out horizontally button in bar and use the hand to drag to the place)
- Add other plots which might be interesting (right click add PV)
- Block as `IN::CS:SB:
- Other PVs
IN:<instrument>:<IOC>:PV.VAL
(these must be marked in the DB file withinfo(archive, "VAL")
)
- Good blocks to choose might be SP and SP:RBVs in this case ploting the SP and SP:RBV showed that the SP had not be set in the machine. So run control hadn't been able to progress and it had hung.